Estimating spatial resolution of in vivo MR images using spatial modulation of magnetisation

Research output: Contribution to journalArticle

3 Citations (Scopus)

Abstract

Spatial Modulation of Magnetization is shown to provide a means of estimating perceived spatial resolution directly in vivo. On the first magnetic resonance system tested, resolution in conventional spin echo images was found to be stability limited in the phase encoding direction and voxel limited (via the Nyquist sampling theorem) in the frequency encoding direction both in vitro and in vivo. As the voxel size approaches half the stripe separation, fringes of resolved and unresolved stripes are formed across the image. This phenomenon is explained and described mathematically. On a second magnetic resonance scanner, resolution in the phase encoding direction of fast spin echo images with centrically ordered phase encoding is shown to be voxel limited in substances with long T2, with poorer resolution in substances with short T2. Resolution in fast spin echo images with linearly ordered phase encoding was shown to be voxel limited in the phase encoding direction.

Original languageEnglish
Pages (from-to)37-44
Number of pages8
JournalMagnetic Resonance Imaging
Volume16
Issue number1
DOIs
Publication statusPublished - Jan 1998

Fingerprint

Magnetic resonance
Magnetization
Modulation
Magnetic Resonance Spectroscopy
Sampling
Direction compound

Keywords

  • performance assessment
  • SPAMM
  • Magnetic resonance

Cite this

Estimating spatial resolution of in vivo MR images using spatial modulation of magnetisation. / Wayte, S C; Redpath, Thomas William.

In: Magnetic Resonance Imaging, Vol. 16, No. 1, 01.1998, p. 37-44.

Research output: Contribution to journalArticle

@article{5dc45940b39b43e68b1153c1b5004947,
title = "Estimating spatial resolution of in vivo MR images using spatial modulation of magnetisation",
abstract = "Spatial Modulation of Magnetization is shown to provide a means of estimating perceived spatial resolution directly in vivo. On the first magnetic resonance system tested, resolution in conventional spin echo images was found to be stability limited in the phase encoding direction and voxel limited (via the Nyquist sampling theorem) in the frequency encoding direction both in vitro and in vivo. As the voxel size approaches half the stripe separation, fringes of resolved and unresolved stripes are formed across the image. This phenomenon is explained and described mathematically. On a second magnetic resonance scanner, resolution in the phase encoding direction of fast spin echo images with centrically ordered phase encoding is shown to be voxel limited in substances with long T2, with poorer resolution in substances with short T2. Resolution in fast spin echo images with linearly ordered phase encoding was shown to be voxel limited in the phase encoding direction.",
keywords = "performance assessment, SPAMM, Magnetic resonance",
author = "Wayte, {S C} and Redpath, {Thomas William}",
year = "1998",
month = "1",
doi = "10.1016/S0730-725X(97)00222-1",
language = "English",
volume = "16",
pages = "37--44",
journal = "Magnetic Resonance Imaging",
issn = "0730-725X",
publisher = "Elsevier",
number = "1",

}

TY - JOUR

T1 - Estimating spatial resolution of in vivo MR images using spatial modulation of magnetisation

AU - Wayte, S C

AU - Redpath, Thomas William

PY - 1998/1

Y1 - 1998/1

N2 - Spatial Modulation of Magnetization is shown to provide a means of estimating perceived spatial resolution directly in vivo. On the first magnetic resonance system tested, resolution in conventional spin echo images was found to be stability limited in the phase encoding direction and voxel limited (via the Nyquist sampling theorem) in the frequency encoding direction both in vitro and in vivo. As the voxel size approaches half the stripe separation, fringes of resolved and unresolved stripes are formed across the image. This phenomenon is explained and described mathematically. On a second magnetic resonance scanner, resolution in the phase encoding direction of fast spin echo images with centrically ordered phase encoding is shown to be voxel limited in substances with long T2, with poorer resolution in substances with short T2. Resolution in fast spin echo images with linearly ordered phase encoding was shown to be voxel limited in the phase encoding direction.

AB - Spatial Modulation of Magnetization is shown to provide a means of estimating perceived spatial resolution directly in vivo. On the first magnetic resonance system tested, resolution in conventional spin echo images was found to be stability limited in the phase encoding direction and voxel limited (via the Nyquist sampling theorem) in the frequency encoding direction both in vitro and in vivo. As the voxel size approaches half the stripe separation, fringes of resolved and unresolved stripes are formed across the image. This phenomenon is explained and described mathematically. On a second magnetic resonance scanner, resolution in the phase encoding direction of fast spin echo images with centrically ordered phase encoding is shown to be voxel limited in substances with long T2, with poorer resolution in substances with short T2. Resolution in fast spin echo images with linearly ordered phase encoding was shown to be voxel limited in the phase encoding direction.

KW - performance assessment

KW - SPAMM

KW - Magnetic resonance

U2 - 10.1016/S0730-725X(97)00222-1

DO - 10.1016/S0730-725X(97)00222-1

M3 - Article

VL - 16

SP - 37

EP - 44

JO - Magnetic Resonance Imaging

JF - Magnetic Resonance Imaging

SN - 0730-725X

IS - 1

ER -