Yielding of the clamped-wire system in the Ilizarov external fixator

Michael Andrew Watson, K. J. Matthias, David William Laurence Hukins, N. Maffulli

    Research output: Contribution to journalArticle

    22 Citations (Scopus)

    Abstract

    This study demonstrates that the clamped-wire system used to suspend bones within an Ilizarov external fixator yields when the perpendicular load exceeds 50 N per wire. Cyclic loading was applied to tensioned wires clamped within an Ilizarov ring component, with steadily increasing load amplitude. Wires were tested at four initial tension settings. The amount of energy lost within the clamped-wire system per load cycle was calculated for every test. The results showed that there was a consistent trend to increasing non-recoverable energy loss per load cycle when peak loads exceed 50 N for all initial tension settings. A finite element (FE) model replicating the experimental conditions was performed to investigate the levels of stress within the loaded wires. The FE model analyses showed that high stresses were generated in the wires close to the clamping sites, and that the stress levels could reasonably be expected to exceed the material yield stress when loaded to about 55 N, for all initial tension settings. The results show that material yield, accompanied by some wire slippage through the clamps, is responsible for system yield, in agreement with previous studies. Although the initial wire tension has an appreciable effect on the wire stiffness, it did not affect the elastic load range of the clamped-wire system. To prevent yield of the clamped-wire system in practice, the fixator should be assembled with sufficient wires to ensure that the load transmitted to each wire by the patient does not exceed 50 N.

    Original languageEnglish
    Pages (from-to)367-374
    Number of pages7
    JournalProceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine
    Volume217
    Issue number5
    DOIs
    Publication statusPublished - 2003

    Keywords

    • Ilizarov
    • external fixation
    • wires
    • yielding
    • tension
    • MECHANICAL PERFORMANCE
    • FRACTURES
    • TENSION

    Cite this

    Yielding of the clamped-wire system in the Ilizarov external fixator. / Watson, Michael Andrew; Matthias, K. J.; Hukins, David William Laurence; Maffulli, N.

    In: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine, Vol. 217, No. 5, 2003, p. 367-374.

    Research output: Contribution to journalArticle

    Watson, Michael Andrew ; Matthias, K. J. ; Hukins, David William Laurence ; Maffulli, N. / Yielding of the clamped-wire system in the Ilizarov external fixator. In: Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine. 2003 ; Vol. 217, No. 5. pp. 367-374.
    @article{61e0efcd65f5495cb247fcaaba95640c,
    title = "Yielding of the clamped-wire system in the Ilizarov external fixator",
    abstract = "This study demonstrates that the clamped-wire system used to suspend bones within an Ilizarov external fixator yields when the perpendicular load exceeds 50 N per wire. Cyclic loading was applied to tensioned wires clamped within an Ilizarov ring component, with steadily increasing load amplitude. Wires were tested at four initial tension settings. The amount of energy lost within the clamped-wire system per load cycle was calculated for every test. The results showed that there was a consistent trend to increasing non-recoverable energy loss per load cycle when peak loads exceed 50 N for all initial tension settings. A finite element (FE) model replicating the experimental conditions was performed to investigate the levels of stress within the loaded wires. The FE model analyses showed that high stresses were generated in the wires close to the clamping sites, and that the stress levels could reasonably be expected to exceed the material yield stress when loaded to about 55 N, for all initial tension settings. The results show that material yield, accompanied by some wire slippage through the clamps, is responsible for system yield, in agreement with previous studies. Although the initial wire tension has an appreciable effect on the wire stiffness, it did not affect the elastic load range of the clamped-wire system. To prevent yield of the clamped-wire system in practice, the fixator should be assembled with sufficient wires to ensure that the load transmitted to each wire by the patient does not exceed 50 N.",
    keywords = "Ilizarov, external fixation, wires, yielding, tension, MECHANICAL PERFORMANCE, FRACTURES, TENSION",
    author = "Watson, {Michael Andrew} and Matthias, {K. J.} and Hukins, {David William Laurence} and N. Maffulli",
    year = "2003",
    doi = "10.1243/095441103770802531",
    language = "English",
    volume = "217",
    pages = "367--374",
    journal = "Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine",
    issn = "0954-4119",
    publisher = "SAGE Publications Ltd",
    number = "5",

    }

    TY - JOUR

    T1 - Yielding of the clamped-wire system in the Ilizarov external fixator

    AU - Watson, Michael Andrew

    AU - Matthias, K. J.

    AU - Hukins, David William Laurence

    AU - Maffulli, N.

    PY - 2003

    Y1 - 2003

    N2 - This study demonstrates that the clamped-wire system used to suspend bones within an Ilizarov external fixator yields when the perpendicular load exceeds 50 N per wire. Cyclic loading was applied to tensioned wires clamped within an Ilizarov ring component, with steadily increasing load amplitude. Wires were tested at four initial tension settings. The amount of energy lost within the clamped-wire system per load cycle was calculated for every test. The results showed that there was a consistent trend to increasing non-recoverable energy loss per load cycle when peak loads exceed 50 N for all initial tension settings. A finite element (FE) model replicating the experimental conditions was performed to investigate the levels of stress within the loaded wires. The FE model analyses showed that high stresses were generated in the wires close to the clamping sites, and that the stress levels could reasonably be expected to exceed the material yield stress when loaded to about 55 N, for all initial tension settings. The results show that material yield, accompanied by some wire slippage through the clamps, is responsible for system yield, in agreement with previous studies. Although the initial wire tension has an appreciable effect on the wire stiffness, it did not affect the elastic load range of the clamped-wire system. To prevent yield of the clamped-wire system in practice, the fixator should be assembled with sufficient wires to ensure that the load transmitted to each wire by the patient does not exceed 50 N.

    AB - This study demonstrates that the clamped-wire system used to suspend bones within an Ilizarov external fixator yields when the perpendicular load exceeds 50 N per wire. Cyclic loading was applied to tensioned wires clamped within an Ilizarov ring component, with steadily increasing load amplitude. Wires were tested at four initial tension settings. The amount of energy lost within the clamped-wire system per load cycle was calculated for every test. The results showed that there was a consistent trend to increasing non-recoverable energy loss per load cycle when peak loads exceed 50 N for all initial tension settings. A finite element (FE) model replicating the experimental conditions was performed to investigate the levels of stress within the loaded wires. The FE model analyses showed that high stresses were generated in the wires close to the clamping sites, and that the stress levels could reasonably be expected to exceed the material yield stress when loaded to about 55 N, for all initial tension settings. The results show that material yield, accompanied by some wire slippage through the clamps, is responsible for system yield, in agreement with previous studies. Although the initial wire tension has an appreciable effect on the wire stiffness, it did not affect the elastic load range of the clamped-wire system. To prevent yield of the clamped-wire system in practice, the fixator should be assembled with sufficient wires to ensure that the load transmitted to each wire by the patient does not exceed 50 N.

    KW - Ilizarov

    KW - external fixation

    KW - wires

    KW - yielding

    KW - tension

    KW - MECHANICAL PERFORMANCE

    KW - FRACTURES

    KW - TENSION

    U2 - 10.1243/095441103770802531

    DO - 10.1243/095441103770802531

    M3 - Article

    VL - 217

    SP - 367

    EP - 374

    JO - Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine

    JF - Proceedings of the Institution of Mechanical Engineers. Part H, Journal of Engineering in Medicine

    SN - 0954-4119

    IS - 5

    ER -